To Test the ZOLL R-Series Defibrillator you will need the following:
- ZOLL R-Series defibrillator
- Your programmed and configured wireless card
- To be within range of the wireless network the card is configured to connect to
To Test the Wireless Transmission:
- Turn the Defibrillator to ‘ON’ (for R-Series BLS) / ‘Monitor Mode’ (for R-Series ALS)
- R-Series BLS Only: Press the ‘Manual Mode’ softkey and confirm
- WAIT – after a few seconds the “Report Data” softkey will appear
- Press ‘Report Data’
- Press ‘Transfer Mode’
- To transfer clinical files (for CaseReview customers) press ‘Report to Wi-Fi’
- To transfer device check history (for Clock Sync or DeviceDashboard customers) press ‘Defib History to Wi-Fi”
- The Defibrillator should show this order of events:
- INITIALIZING NETWORK
- CONNECTING TO NETWORK
- WAITING FOR SERVER RESPONSE: Progress bar should be at 100%
- TRANSFER COMPLETE: Full Disclosure Was Sent (If you pressed “Report to Wi-Fi)
OR
- TRANSFER COMPLETE: Device Check Was Sent (if you pressed “Defib History to Wi-Fi”)
- If the transfer failed for any reason, retry. If you receive errors 4+ times in a row, contact technical support for assistance: https://www.zoll.com/contact/customer-service