Global Change
AED 3 – Configure Wireless Connections | ZOLL Medical

AED 3 – Configure Wireless Connections

To begin configuring wireless connections, navigate to the Wi-Fi Settings menu:


From the “Advanced Configuration” menu, press the “Wi-Fi” icon to enter the “Wi-Fi Settings” menu.

** Note for Configuration: Press Ok/Save after all configuration page changes to ensure settings are retained. Pressing the back arrow instead may discard your changes.

  1. Network Settings
    1. Select DHCP or Static IP
      1. If you select Static IP, each defib will have its own assigned IP address on the network. (This is less common and will require unique configurations for each AED).
      2. Static IP Settings will need to be provided by your local IT department.
  2. Access Point List
    1. Add/Edit Access Points
      1. SSID: Enter the wireless network’s name here.
      2. Authentication:
        • PSK – will only require a Password (Pre-Shared Key)
        • PEAP – will require a Username and Password
        • TLS – will require a Username, Password, and Client Certificate
        • The Client Certificate must be a .pfx (binary) file. Insert a USB 2.0 compliant flash drive into the AED 3 then select ‘Client Certificate’ on this configuration menu and choose your certificate from the list. For additional information, refer to the AED 3 Administrator’s guide or contact ZOLL Technical Support: https://www.zoll.com/contact/customer-service
    2. Press ‘Save’
  3. Clinical Archive Settings – (RescueNet CaseReview Customers)
    1. Server Settings:
      1. ‘Address’ refers to the URL or IP address. For customers transmitting directly to ZOLL Online, choose ‘URL’ and enter:
        • US Cloud Customers: dxsvc.zollonline.com
        • Canadian Cloud Customers: dxsvc.ca.zollonline.com
        • In-Premise customers: enter the URL (recommended) or IP address as provided by your IT department.
      2. ‘Port’ – default is 443
      3. ‘User ID’ and ‘Passcode’
        • In-Premise customers will have had their User ID and Passcode configured during the In-Premise software installation process. If this information is not known, contact ZOLL technical support for assistance: https://www.zoll.com/contact/customer-service
        • For customers submitting directly to ZOLL Online, you will enter the DAK “Alternate ID” and “Password” here.
    2. Select Access Point: choose the access point you created previously.
    3. Case Push Enabled: On/Off (On - must be selected for clinical data to transfer)
    4. Press ‘Save’
  4. Device History Settings (RescueNet DeviceDashboard or DefibrillatorDashboard customers)
    1. Server Settings:
      1. ‘Address’ refers to the URL or IP address. For customers transmitting directly to ZOLL Online, choose ‘URL’ and enter:
        • US Cloud Customers: dxsvc.zollonline.com
        • Canadian Cloud Customers: dxsvc.ca.zollonline.com
        • In-Premise customers: enter the URL (recommended) or IP address as provided by your IT department.
      2. ‘Port’ – default is 443
      3. ‘User ID’ and ‘Passcode’
        • In-Premise customers will have had their User ID and Passcode configured during the In-Premise software installation process. If this information is not known, contact ZOLL technical support for assistance: https://www.zoll.com/contact/customer-service
        • For customers submitting directly to ZOLL Online, you will enter the DAK “Alternate ID” and “Password” here.
    2. Select Access Point: choose the access point you created previously.
    3. Self-Test Settings:
      1. Auto-Self Test Report – On/Off (On is mandatory for DeviceDashboard and DefibDashboard customers).
      2. Self-Test Interval: Choose either 1 (daily) or 7 (weekly) self-tests – 7 Days is recommended to prolong battery life.
      3. Custom Settings: (On/Off) – If ‘On’, you can choose a specific time of day for the self-test to complete.
    4. Press ‘Save’