To begin configuring wireless connections, navigate to the Wi-Fi Settings menu:

From the “Advanced Configuration” menu, press the “Wi-Fi” icon to enter the “Wi-Fi Settings” menu.
** Note for Configuration: Press Ok/Save after all configuration page changes to ensure settings are retained. Pressing the back arrow instead may discard your changes.
- Network Settings
- Select DHCP or Static IP
- If you select Static IP, each defib will have its own assigned IP address on the network. (This is less common and will require unique configurations for each AED).
- Static IP Settings will need to be provided by your local IT department.
- Access Point List
- Add/Edit Access Points
- SSID: Enter the wireless network’s name here.
- Authentication:
- PSK – will only require a Password (Pre-Shared Key)
- PEAP – will require a Username and Password
- TLS – will require a Username, Password, and Client Certificate
- The Client Certificate must be a .pfx (binary) file. Insert a USB 2.0 compliant flash drive into the AED 3 then select ‘Client Certificate’ on this configuration menu and choose your certificate from the list. For additional information, refer to the AED 3 Administrator’s guide or contact ZOLL Technical Support: https://www.zoll.com/contact/customer-service
- Press ‘Save’
- Clinical Archive Settings – (RescueNet CaseReview Customers)
- Server Settings:
- ‘Address’ refers to the URL or IP address. For customers transmitting directly to ZOLL Online, choose ‘URL’ and enter:
- US Cloud Customers: dxsvc.zollonline.com
- Canadian Cloud Customers: dxsvc.ca.zollonline.com
- In-Premise customers: enter the URL (recommended) or IP address as provided by your IT department.
- ‘Port’ – default is 443
- ‘User ID’ and ‘Passcode’
- In-Premise customers will have had their User ID and Passcode configured during the In-Premise software installation process. If this information is not known, contact ZOLL technical support for assistance: https://www.zoll.com/contact/customer-service
- For customers submitting directly to ZOLL Online, you will enter the DAK “Alternate ID” and “Password” here.
- Select Access Point: choose the access point you created previously.
- Case Push Enabled: On/Off (On - must be selected for clinical data to transfer)
- Press ‘Save’
- Device History Settings (RescueNet DeviceDashboard or DefibrillatorDashboard customers)
- Server Settings:
- ‘Address’ refers to the URL or IP address. For customers transmitting directly to ZOLL Online, choose ‘URL’ and enter:
- US Cloud Customers: dxsvc.zollonline.com
- Canadian Cloud Customers: dxsvc.ca.zollonline.com
- In-Premise customers: enter the URL (recommended) or IP address as provided by your IT department.
- ‘Port’ – default is 443
- ‘User ID’ and ‘Passcode’
- In-Premise customers will have had their User ID and Passcode configured during the In-Premise software installation process. If this information is not known, contact ZOLL technical support for assistance: https://www.zoll.com/contact/customer-service
- For customers submitting directly to ZOLL Online, you will enter the DAK “Alternate ID” and “Password” here.
- Select Access Point: choose the access point you created previously.
- Self-Test Settings:
- Auto-Self Test Report – On/Off (On is mandatory for DeviceDashboard and DefibDashboard customers).
- Self-Test Interval: Choose either 1 (daily) or 7 (weekly) self-tests – 7 Days is recommended to prolong battery life.
- Custom Settings: (On/Off) – If ‘On’, you can choose a specific time of day for the self-test to complete.
- Press ‘Save’