X-Series – Recommended Data Transfer Settings | ZOLL Medical

X-Series – Recommended Data Transfer Settings


The X-Series has all data transfer settings configured correctly by default with only one exception that applies to DeviceDashboard/DefibDashboard customers:

  1. Enter the Supervisor menu
  2. Open “Communications”
  3. Select “Readiness Test”
  4. Select “Configure”
  5. Toggle “Auto Transmit Results” to “Enable” to have the X-Series automatically transmit the readiness test results following a manual 30-Joule test