Corrective Action for the AED Plus

Previously ZOLL had recommended users either replace batteries every three years, or optionally upgrade their software to the current version (5.32) to detect battery problems within the maximum five-year standby life. We are now recommending that ALL users of the AED Plus upgrade their software to the current version (5.32).  

Our original offer to provide a sleeve of ten Lithium 123A batteries for $39.95 (including shipping) is still in effect.  Users who submit a valid serial number may still purchase batteries at the reduced price from this website.