On February 12, 2009, ZOLL® recommended that all AED Plus® users whose units have serial numbers with the last six digits below 200000 replace the 123A Lithium batteries in their unit every three years to prevent potential battery failures in attempted rescues. In addition, ZOLL recommends that a label be applied to all such units indicating when batteries need to be replaced. This web site provides everything ZOLL AED Plus users need to take corrective action.
Find out more:
Customer Notification Letter
Frequently Asked Questions
Labeling Instructions
|
|

Look at the unique serial number printed on the bottom of your AED Plus. If the last six digits are UNDER 200000, then your AED Plus IS subject to the Corrective Action. |
|

If your current batteries were installed more than 3 years ago, you should replace them.
Order 10 replacement batteries with a replacement label from ZOLL at a special price.
If your batteries were installed less than 3 years ago, you should attach a Battery Replacement Label to the unit. Order labels at no charge. |
|
|

|
Rather not replace batteries? You may not have to. Upgrade your AED Plus with new software that can detect if and when your batteries need replacement. |
|
|

Upgrade your device.
|
|
|
|
|
|